Section 7: Capturing and analyzing waveforms
Model DMM7510 7½ Digit Multimeter Application Manual
7-12
DMM7510-904-01 Rev. D March 2021
Figure 18: Inductor current linearity test connections
Run the test using the front panel
Running the inductor current linearity test from the front panel:
1. Press the
POWER
button on the front panel to turn on the instrument.
2. On the FUNCTIONS swipe screen, select
Digi I
to select the digitize current function.
3. Swipe to the SETTINGS swipe screen.
4. Set the Sample Rate to
500,000
.
5. Set the Count to
50
.
6. Press the
MENU
key.
7. Select
Graph
to view the graph.