Commands listed in alphabetical order
PERIOD|PHASE|PHASENOISE|PJ|PK2Pk| POVERSHOOT|PWIDTH|QFACTOR|
RISESLEWRATE|RISETIME|RJ|RJDIRAC|RMS|
SETUP|SKEW|SRJ|SSCFREQDEV|SSCMODRATE|TIE|TIMEOUTSIDELEVEL|
TJBER|TNTRATIO|TOP|UNITINTERVAL|VDIFFXOVR|WIDTH|WIDTHBER}
Arguments
Arguments are available measurements.
Examples
MEASUREMENT:ADDMEAS FREQUENCY
adds a frequency measurement.
MEASUrement:ADDNew (No Query Form)
This command adds the speci
fi
ed measurement.
Group
Measurement
Syntax
MEASUrement:ADDNew
"QString"
Arguments
"QString"
is the measurement to add. The argument is of the form
"MEAS<NR1>" where NR1
≥
1.
Examples
MEASUrement:ADDNew "MEAS11"
adds measurement 11.
MEASUrement:ANNOTate
This command sets or queries the annotation state for measurements.
Group
Measurement
Syntax
MEASUrement:ANNOTate {OFF|AUTO}
MEASUrement:ANNOTate?
Arguments
OFF
turns off measurement annotations.
AUTO
turns on visible measurement annotations.
Examples
MEASUREMENT:ANNOTATE OFF
turns off measurement annotations.
MEASUREMENT:ANNOTATE?
might return
:MEASUREMENT:ANNOTATION:STATE
OFF
, indicating that no measurement annotations are active.
MSO54, MSO56, MSO58, MSO58LP Programmer
2-381
Summary of Contents for MSO54
Page 1: ...xx 5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual P077130502 077 1305 02 ...
Page 2: ......
Page 3: ...5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual xx www tek com 077 1305 02 ...
Page 8: ...Preface iv MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 26: ...Command Syntax 2 10 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 936: ...Status and Events 3 20 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 937: ...Appendix A Character Set MSO54 MSO56 MSO58 MSO58LP Programmer A 1 ...
Page 938: ...Appendix A Character Set A 2 MSO54 MSO56 MSO58 MSO58LP Programmer ...