Commands listed in alphabetical order
TRIGger:{A|B}:BUS:B<x>:MIL1553B:STATus:ADDRess:QUALi
fi
er
This command sets or queries the quali
fi
er to be used when triggering on status
word addresses for a MIL-STD-1553 bus. The bus number is speci
fi
ed by x. The
trigger condition must be set to STATus.
Conditions
Requires the SR-AERO Triggering and Analysis application.
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:MIL1553B:STATus:
ADDRess:QUALifier
{EQual|UNEQual|LESSthan|MOREthan
|LESSEQual|MOREEQual|INrange|OUTrange}
TRIGger:{A|B}:BUS:B<x>:MIL1553B:STATus:ADDRess:QUALifier?
Arguments
Arguments are the available address quali
fi
ers.
Examples
TRIGGER:A:BUS:B1:MIL1553B:STATUS:ADDRESS:QUALIFIER LESSTHAN
sets the address quali
fi
er to less than.
TRIGGER:A:BUS:B1:MIL1553B:STATUS:ADDRESS:QUALIFIER?
might return
EQUAL, indicating that the address quali
fi
er is set to equal.
TRIGger:{A|B}:BUS:B<x>:MIL1553B:STATus:ADDRess:VALue
This command sets or queries the low value when triggering on status word
addresses for a MIL-STD-1553 bus. The bus number is speci
fi
ed by x. The trigger
condition must be set to STATus.
Conditions
Requires the SR-AERO Triggering and Analysis application.
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:MIL1553B:STATus:ADDRess:VALue
<QString>
TRIGger:{A|B}:BUS:B<x>:MIL1553B:STATus:ADDRess:VALue?
Arguments
<QString>
is the address value.
2-822
MSO54, MSO56, MSO58, MSO58LP Programmer
Summary of Contents for MSO54
Page 1: ...xx 5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual P077130502 077 1305 02 ...
Page 2: ......
Page 3: ...5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual xx www tek com 077 1305 02 ...
Page 8: ...Preface iv MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 26: ...Command Syntax 2 10 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 936: ...Status and Events 3 20 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 937: ...Appendix A Character Set MSO54 MSO56 MSO58 MSO58LP Programmer A 1 ...
Page 938: ...Appendix A Character Set A 2 MSO54 MSO56 MSO58 MSO58LP Programmer ...