Commands listed in alphabetical order
TRIGger:{A|B}:BUS:B<x>:USB:DATa:VALue
This command speci
fi
es the binary data string to be used when triggering on a
USB trigger.
Conditions
Requires the SR-USB2 Serial Triggering and Analysis Application.
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:USB:DATa:VALue <QString>
TRIGger:{A|B}:BUS:B<x>:USB:DATa:VALue?
Related Commands
TRIGger:{A|B}:BUS:B<x>:USB:CONDition
Arguments
<QString>
within the range 00000000 to 11111111 (00 hex to FF hex).
Examples
TRIGGER:A:BUS:B1:USB:DATA:VALUE "00001000"
sets the binary address
to 00001000 (08 hex).
TRIGGER:A:BUS:B1:USB:DATA:VALUE?
might return
:TRIGGER:A:BUS:B1:USB:DATA:VALUE "01000000"
, which
indicates that the binary address is 0100000 (40 hex).
TRIGger:{A|B}:BUS:B<x>:USB:ENDPoint:VALue
This command speci
fi
es the binary endpoint string to be used for the USB trigger.
The bus number is speci
fi
ed by <x>.
Conditions
Requires the SR-USB2 Serial Triggering and Analysis Application.
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:USB:ENDPoint:VALue <QString>
TRIGger:{A|B}:BUS:B<x>:USB:ENDPoint:VALue?
Related Commands
TRIGger:{A|B}:BUS:B<x>:USB:CONDition
Arguments
<QString>
within the range 0000 to 1111 (00 hex to 0F hex).
2-840
MSO54, MSO56, MSO58, MSO58LP Programmer
Summary of Contents for MSO54
Page 1: ...xx 5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual P077130502 077 1305 02 ...
Page 2: ......
Page 3: ...5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual xx www tek com 077 1305 02 ...
Page 8: ...Preface iv MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 26: ...Command Syntax 2 10 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 936: ...Status and Events 3 20 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 937: ...Appendix A Character Set MSO54 MSO56 MSO58 MSO58LP Programmer A 1 ...
Page 938: ...Appendix A Character Set A 2 MSO54 MSO56 MSO58 MSO58LP Programmer ...