background image

Fig.  2-2.  Current flow in a conductor. 

High  Currents 

When  measuring  high  currents,  do  not  leave  the  current 

probe  clamped  around  the  conductor  while  disconnecting 
the  probe  cable  from  the  termination.  With  the  probe  cable 
unterminated  in  this  condition,  a  high  voltage  is  developed 

in  the  secondary  winding  which  may  damage  the  current 

probe transformer. 

Increasing  the  Sensitivity 

The  current  sensitivity  of  the  P6021  Current  Probe  and 

Termination  can  be  increased  by  increasing  the  number  of 
turns  passing  through  the  core  of  the  probe.  For  example, 
if  the  conductor  is  looped  through  the  probe  two  times,  a 
two-turn  primary  winding  is  formed,  increasing  the  second­
ary  current  by  a  factor  of  two.  (The  ratio  of  current  in  a 
transformer  is  inversely  proportional  to  the  turns  ratio.)  With 
the  P6021  Termination  switch  set  to  2 mA/mY  and  the  os­
cilloscope  vertical  deflection  factor  set  to  10 mY /div,  the 
deflection  factor  is  actually  reduced  by  a  factor  of  two  to 

10  mA/div. 

Remember,  however,  that  the  impedance  reflected  into 

the  primary  (circuit  being  measured)  from  the  secondary 
(probe  winding)  varies  as  the  square  of  the  primary  turns. 
When  observing  high-frequency  current  waveforms  or  fast­
rise  pulses,  the  inductance  added  to  the  primary  circuit  by 
the additional turns may be significant. 

Insertion  Impedance 

The  insertion  impedance  of  the  current  probe  is  the  equiv­

alent  circuit  which  is  placed  in  the  circuit  under  test  when 

Operating  Instructions-P6021 

2.8 

pH 

1 .7 nH 

.004 

Fig.  2-3.  Insertion  impedance  of  the  P6021  Current  Probe. 

the  probe  is  clamped  around  a  conductor.  When  observing 
fast-rise  signals,  this  should  be  taken  into  consideration. 
Fig.  2-3  illustrates  the  approximate  insertion  impedance  of 
the P602 1 Current Probe. 

Probe  Shielding 

The  current  probe  is  shielded  to  minimize  the  effect  of 

external  magnetic  fields.  However,  strong  fields  may  in­
terfere  with  a  current  signal  being  measured.  If  you  suspect 

that  an  external  field  is  interfering  with  your  measurement, 
remove  the  probe  from  the  conductor  and  place  it  in  the 
vicinity  of  the  original  measurement.  If  you  obtain  appreci­
able  deflection,  attempt  to  measure  the  conductor  current  at 
another  point,  away  from  the  magnetic  field  source. 

If  current  measurements  must  be  made  in  the  presence  of 

a  strong  external  field,  the  external  field  interference  may 
be  minimized  by  the  use  of  two  current  probes  and  a  dif­

ferential-input  oscilloscope.  Both  current  probes  must  be 

connected  to  the  oscilloscope  inputs  in  the  same  manner, 
using  P6021 Termination. 

With  both  probes  connected  to  a  differential-input  oscil­

loscope,  clamp  one  probe  around  the  conductor  in  which 
the  current  is  to  be  measured,  and  place  the  other  probe 

near  the  first,  with  the  slider  closed.  By  setting  the  oscillo­
scope  controls  for  common-mode  rejection,  the  undesirable 

current  signal  induced  in  one  probe  can  be  minimized  by  the 

induced  current  in  a  second  probe.  Adjust  the  positions  of 
the  probes  for  best  results.  Complete  cancellation  of  the 

undesirable  signal  may  be  difficult  to  obtain  due  to  probe 

and  termination  differences. 

2-2 

Summary of Contents for P6021

Page 1: ...IMfJuu1iD1t MIJIWlIl P6021 CURRENT PROBE AND TERMINATION Tektronix Inc P O Box 500 Beaverton Oregon 97077 Phone 644 161 Cables Tektronix 070 0947 00 First Printing MAR 1969 Revised MAY 1981 ...

Page 2: ...ion of Tektronix Inc Products of Tektronix Inc and its subsidiaries are covered by U S and foreign patents and or pending patents TEKTRONIX TEK SCOPE MOBILE and are registered trademarks of Tektronix Inc TELEQUIP MENT is a registered trademark of Tektronix U K Limited Printed in U S A Specification and price change privileges are reserved ...

Page 3: ...Verso Filler Page z z ...

Page 4: ...Fig 1 1 P6021 Current Probe and Termination ...

Page 5: ... Aberrations Probe and Termination only 2 mA mV and 4 4 or less not to exceed 10 mA mV 5 peak to peak within 50 nanoseconds of step 1 1 or less total not to exceed 2 peak to peak thereafter Risetime Probe and Termination only 2 mA mV 4 5 nanoseconds or less 10 mA mV 4 nanoseconds or less Characteristic Risetime with 580 Series test oscillo scope 2 mA mV 10 mA mV Tilt 2 mA mV 10 mA mV Sensitivity A...

Page 6: ...depending upon the amplitude of the signal The overall deflection factor including the probe and termination may be found quickly by multiplying the slide switch position by the oscil loscope deflection factor The following is an example Termination switch setting 1 0 mA mV Volts div switch setting 20 mV div 1 0 mA 20 mV 200 mA mV X div div Probe Slider The thumb controlled probe slider opens the ...

Page 7: ...ing Instructions P6021 2 8 pH 1 7 nH 004 n Fig 2 3 Insertion impedance of the P6021 Current Probe the probe is clamped around a conductor When observing fast rise signals this should be taken into consideration Fig 2 3 illustrates the approximate insertion impedance of the P602 1Current Probe Probe Shielding The current probe is shielded to minimize the effect of external magnetic fields However s...

Page 8: ...fference in level between the pulses induced in the paralled windings of the secondary and matches the balanced probe winding to the cable 3 1 As indicated on the probe body the turns ratio of the P6021 Curent Probe is 125 1 This refers to the number of windings in the secondary of the probe transformer The two turns ratios manufactured by Tektronix Inc are 50 1 and 125 1 Due to the comparatively ...

Page 9: ...r in the slider 7 Remove the two Phillips head screws which secure the spring retainer holder to the probe body Remove the holder 8 Lift the circuit board transformer and cable out of the probe body as a unit The transformer may be unplugged from the circuit board The P6021 Current Probe is shown disassembled in the Mechanical Parts List section 9 Reassemble the probe reversing the above procedure...

Page 10: ...ust ment is made For best overall instrument performance make each adustment to the exact setting even if the CHECK is within the allowable tolerance EQUIPMENT REQUIRED General The following items are required for calibration of the P6021 Current Probe and Termination Specifications given 5 1 are the minimum necessary for accurate calibration All test equipment is assumed to be correctly calibrate...

Page 11: ...ring first 50 nanoseconds 1 1 total of 2 peak to peak thereafter 9 Foot Probe Aberrations within 4 4 total of 5 peak to peak during first 50 nanoseconds 1 1 total of 2 peak to peak thereafter o 3 Check Risetime REQUIREMENT 1 0 mA mV risetime of 6 5 nano seconds 2 mA mV risetime of 6 7 nan seconds PERFORMANCE Correct ___ incorrect list ex ceptions __ _ o 4 Check Sensitivity REQUIREMENT 2 mA mV sens...

Page 12: ...k must be constructed to match the 62 S ohm probe coble to the 50 ohm termination Using a GR insertion unit construct the network shown in Fig 5 2 Mark the insertion unit for direction and place the cover over the openings H From current probe GR InSertiOn unit To test oscilloscope Fig 5 2 Impedance matching network constructed in GR insertion unit Remove the top cover only from the P6022 Terminat...

Page 13: ...ation slide switch to the 2 mA mY position b Connect the P6021 Probe from the Calibration Fixture to the termination c Adjust the pulse generator output amplitude for four centimeters of deflection Position the top of the display within the CRT viewing area d CHECK 5 Foot Probe First 50 nanoseconds of step for aberrations within 3 3 total of 4 peak to peak 1 1 total of 2 peak to peak after the fir...

Page 14: ...e termination slide switch to the 1 0 mA mV position I CHECK CRT display for four centimeters of deflection 3 not including GR attenuator error See Fig 5 5B 5 Check Low Frequency Response a Connect the banana plug to BNC adopter the BNC male to GR adapter and the Calibration Fixture to the out put of the low frequency signal generator Connect the P6021 Probe to the Calibration Fixture b Set the te...

Page 15: ... 00 283 01 82 00 1 08 0526 00 1 08 0526 00 1 08 0529 00 Capacitors 2 5 9 pF Var 51 pF Inductors Resistors 50 p H 50 p H 1 6 p H Cer Resistors are fixed composition 1 0 unless otherwise indicated R1 0 R1 2 R1 3 R1 4 T1 T14 REV B JUNE 1975 311 0635 00 31 1 0635 00 317 0821 00 31 7 0121 00 120 0614 00 1 20 0468 00 1 k n Var 1 k n Var 820n 120n Transformers Current Ve W l e W Toroid 6 turns bifilar 40...

Page 16: ... 283 0140 00 1 31 0602 00 131 0106 02 1 08 0525 00 108 0395 00 fixed composition 317 0101 00 321 0077 00 321 0069 00 321 0039 00 321 0038 00 317 0036 00 311 0605 00 260 0723 00 1 0 unless 5 25 pF Var 5 25 pF Var 4 7 pF Connectors Cer Receptacle electrical male Receptacle electrical female Inductors Resistors 140 fLH 64 fLH otherwise indicated 1000 61 90 51 10 24 90 24 30 3 60 2000 Var Switch Slide...

Page 17: ...erio 1 Model No t Description Eff Disc y 1 2 3 4 5 6948 PROBE PACKAGE w termination 6949 PROBE PACKAGE w termination PROBE PACKAGE w termination 9 root probe probe package includes TERMINATION coaxial 6948 PROBE PACKAGE w accessories 6949 PROBE PACKAGE w accessories PROBE PACKAGE w accessories 9 foot probe probe package includes 6948 1 PROBE P6021 6949 1 PROBE P6021 1 PROBE P6201 9 foot 2 CLIP pro...

Page 18: ...on y 1 2 3 4 5 PROBE P6021 PROBE P6021 PROBE P6021 9 foot probe includes 1 BODY probe top 1 BALL metal 0 094 inch OD 1 SLIDE ASSEMBLY upper shell 1 CONTACT electrical 1 T1 TRANSFORMER 1 SPRING helical compression 1 RETAINER return spring 1 HOLDER spring return 2 SCREW 2 56 x 1 4 inch RHS 1 BODY probe bottom 1 BODY probe bottom 1 BODY probe bottom 9 foot probe 1 HOLDER circuit board 6 SOCKET pin co...

Page 19: ...TION Q Serial Model No Eff Disc Y 1 2 J 4 5 TERMINATION P6021 termination includes Description COVER termination top w knob CONNECTOR electrical BNC male FRAME termination CONNECTOR BNC female ASSEMBLY circuit board assembly includes INSULATOR plate BOARD circuit mounting hardware not included w assembly 2 SPACER sleeve 0 087 ID x 0 1 87 inch OD 2 WASHER fiat 0 090 ID x 0 188 inch OD 2 LOCKWASHER ...

Page 20: ... FORMER I I LI4 I 3T J R3b CoRE r 04 TI4 t P2 0 J 20 rl J R2ZJ n f f P39 LIO I I I I 1 _ 50M RIS 820 LI2 I I I I I I 5O u 1 4 CI3 2 I I I 1 ADDED AFTER 1 1 2 49 I C 50 C 2Z 5 25 L 31 t04 H R35 3 1 34 2 4 5 i III I I I II 5 ILLU L_____________L _____ _ I 1 I I I __ J 4 1 P o02 1 PROBE PG02 1 PA551VE TE RMINATIOt l I TYPE P60 2 J CURRENT PROBE TERMINATION 0670 R EV APR 1974 ...

Reviews: