background image

I

P6021 

WARNING 

THE  FOLLOWING  SERVICING  INSTRUCTIONS 
ARE FOR USE BYQUALIFIED PERSONNEL ONLY. 
TO  AVOID  PERSONAL  INJURY,  DO  NOT 

PERFORM  ANY  SERVICING  OTHER  THAN  THAT 

CONTAINED  IN  OPERATING  INSTRUCTIONS 

UNLESS YOU ARE QUALIFIED TO DO SO. 

SECTION 

CIRCUIT  DESCRIPTION 

Introduction 

This  section  of  the  manual  contains  descriptions  of  the 

circuitry  in  the  P6021  Current  Probe  and  Termination.  Com· 
plete  circuit  schematics  are  given  in  the  Diagrams  section. 
Ref'er  to  these  diagrams  throughout  the  following  circuit 

descri ption. 

Current  Probe 

The  P6021  Current  Probe  consists  of 

current  transformer 

mounted  in  the  nose  of  the  case,  an  impedance  matching 
network,  and  a  switch  to  disconnect  the  transformer  shield 
from  ground. 

The  transformer  contains  a  two-section  U-shoped  ferrite 

core.  One  section  is  stationary,  the  other  is  mechanically 
movable  to  permit  closing  the  core  around  the  conductor 

being  measured.  The  conductor  under  test  forms  a  one·turn 
primary  winding  for  the  transformer;  the  windings  around 
the  stationary  portion  of  the  core  form  the  secondary  wind­
ings.  Paralled  windings  in  the  secondary  assure  a  fast  step 

response.  The  circuitry  between  the  transformer  and  the 

coaxial  cable  corrects  any  difference  in  level  between  the 
pulses  induced  in  the  paralled  windings  of  the  secondary, 

and  matches  the  balanced  probe  winding  to  the  cable. 

3-1 

As  indicated  on  the  probe  body,  the  turns  ratio  of  the 

P6021  Curent  Probe  is  125:1 .  This  refers  to  the  number  of 
windings  in  the  secondary  of  the  probe  transformer.  The 
two  turns  ratios  manufactured  by  Tektronix,  Inc.  are  50,1 
and  125,1 .  Due  to  the  comparatively  high  inductance  in 
the  secondary  of  the  1 25:1  probe,  the  P6021  operates  at  a 
lower  frequency  range  than  the  50; 1  probe. 

The  probe  transformer  is  shielded  to  eliminate  interference 

from  outside  signals.  To  eliminate  the  possibility  of  shorting 

this  shield  to  the  conductor  being  measured  when  connecting 

or  removing  the  probe,  slide  switch  SW20  disconnects  the 

ground  from  the  shield  when  the  slide  portion  of  the  probe 

is open. 

Termination 

The  P602 1 Termination  consists  of  a  62.5-ohm  impedance 

matching  network  to  terminate  the  coaxial  cable,  and  a 
voltage  divider  which  is  switched  in  by  a  slide  switch, 

changing  the  sensitivity  by  a  factor  of  five.  With  the  slide 

switch  in  the  2 mA/mV  position,  a  ten  milliampere  current 
signal  in  the  conductor  under  test  induces  a  five  millivolt 
signal  at  the  output  of  the  termination,  when  the  termination 
is  connected  to  a  one-megohm  input  oscilloscope.  In  the 
10 mA/mV  position,  the  signal  is  attenuated  to  produce  a 
one millivolt signal at the output. 

REV 

B  JUN  1980 

Summary of Contents for P6021

Page 1: ...IMfJuu1iD1t MIJIWlIl P6021 CURRENT PROBE AND TERMINATION Tektronix Inc P O Box 500 Beaverton Oregon 97077 Phone 644 161 Cables Tektronix 070 0947 00 First Printing MAR 1969 Revised MAY 1981 ...

Page 2: ...ion of Tektronix Inc Products of Tektronix Inc and its subsidiaries are covered by U S and foreign patents and or pending patents TEKTRONIX TEK SCOPE MOBILE and are registered trademarks of Tektronix Inc TELEQUIP MENT is a registered trademark of Tektronix U K Limited Printed in U S A Specification and price change privileges are reserved ...

Page 3: ...Verso Filler Page z z ...

Page 4: ...Fig 1 1 P6021 Current Probe and Termination ...

Page 5: ... Aberrations Probe and Termination only 2 mA mV and 4 4 or less not to exceed 10 mA mV 5 peak to peak within 50 nanoseconds of step 1 1 or less total not to exceed 2 peak to peak thereafter Risetime Probe and Termination only 2 mA mV 4 5 nanoseconds or less 10 mA mV 4 nanoseconds or less Characteristic Risetime with 580 Series test oscillo scope 2 mA mV 10 mA mV Tilt 2 mA mV 10 mA mV Sensitivity A...

Page 6: ...depending upon the amplitude of the signal The overall deflection factor including the probe and termination may be found quickly by multiplying the slide switch position by the oscil loscope deflection factor The following is an example Termination switch setting 1 0 mA mV Volts div switch setting 20 mV div 1 0 mA 20 mV 200 mA mV X div div Probe Slider The thumb controlled probe slider opens the ...

Page 7: ...ing Instructions P6021 2 8 pH 1 7 nH 004 n Fig 2 3 Insertion impedance of the P6021 Current Probe the probe is clamped around a conductor When observing fast rise signals this should be taken into consideration Fig 2 3 illustrates the approximate insertion impedance of the P602 1Current Probe Probe Shielding The current probe is shielded to minimize the effect of external magnetic fields However s...

Page 8: ...fference in level between the pulses induced in the paralled windings of the secondary and matches the balanced probe winding to the cable 3 1 As indicated on the probe body the turns ratio of the P6021 Curent Probe is 125 1 This refers to the number of windings in the secondary of the probe transformer The two turns ratios manufactured by Tektronix Inc are 50 1 and 125 1 Due to the comparatively ...

Page 9: ...r in the slider 7 Remove the two Phillips head screws which secure the spring retainer holder to the probe body Remove the holder 8 Lift the circuit board transformer and cable out of the probe body as a unit The transformer may be unplugged from the circuit board The P6021 Current Probe is shown disassembled in the Mechanical Parts List section 9 Reassemble the probe reversing the above procedure...

Page 10: ...ust ment is made For best overall instrument performance make each adustment to the exact setting even if the CHECK is within the allowable tolerance EQUIPMENT REQUIRED General The following items are required for calibration of the P6021 Current Probe and Termination Specifications given 5 1 are the minimum necessary for accurate calibration All test equipment is assumed to be correctly calibrate...

Page 11: ...ring first 50 nanoseconds 1 1 total of 2 peak to peak thereafter 9 Foot Probe Aberrations within 4 4 total of 5 peak to peak during first 50 nanoseconds 1 1 total of 2 peak to peak thereafter o 3 Check Risetime REQUIREMENT 1 0 mA mV risetime of 6 5 nano seconds 2 mA mV risetime of 6 7 nan seconds PERFORMANCE Correct ___ incorrect list ex ceptions __ _ o 4 Check Sensitivity REQUIREMENT 2 mA mV sens...

Page 12: ...k must be constructed to match the 62 S ohm probe coble to the 50 ohm termination Using a GR insertion unit construct the network shown in Fig 5 2 Mark the insertion unit for direction and place the cover over the openings H From current probe GR InSertiOn unit To test oscilloscope Fig 5 2 Impedance matching network constructed in GR insertion unit Remove the top cover only from the P6022 Terminat...

Page 13: ...ation slide switch to the 2 mA mY position b Connect the P6021 Probe from the Calibration Fixture to the termination c Adjust the pulse generator output amplitude for four centimeters of deflection Position the top of the display within the CRT viewing area d CHECK 5 Foot Probe First 50 nanoseconds of step for aberrations within 3 3 total of 4 peak to peak 1 1 total of 2 peak to peak after the fir...

Page 14: ...e termination slide switch to the 1 0 mA mV position I CHECK CRT display for four centimeters of deflection 3 not including GR attenuator error See Fig 5 5B 5 Check Low Frequency Response a Connect the banana plug to BNC adopter the BNC male to GR adapter and the Calibration Fixture to the out put of the low frequency signal generator Connect the P6021 Probe to the Calibration Fixture b Set the te...

Page 15: ... 00 283 01 82 00 1 08 0526 00 1 08 0526 00 1 08 0529 00 Capacitors 2 5 9 pF Var 51 pF Inductors Resistors 50 p H 50 p H 1 6 p H Cer Resistors are fixed composition 1 0 unless otherwise indicated R1 0 R1 2 R1 3 R1 4 T1 T14 REV B JUNE 1975 311 0635 00 31 1 0635 00 317 0821 00 31 7 0121 00 120 0614 00 1 20 0468 00 1 k n Var 1 k n Var 820n 120n Transformers Current Ve W l e W Toroid 6 turns bifilar 40...

Page 16: ... 283 0140 00 1 31 0602 00 131 0106 02 1 08 0525 00 108 0395 00 fixed composition 317 0101 00 321 0077 00 321 0069 00 321 0039 00 321 0038 00 317 0036 00 311 0605 00 260 0723 00 1 0 unless 5 25 pF Var 5 25 pF Var 4 7 pF Connectors Cer Receptacle electrical male Receptacle electrical female Inductors Resistors 140 fLH 64 fLH otherwise indicated 1000 61 90 51 10 24 90 24 30 3 60 2000 Var Switch Slide...

Page 17: ...erio 1 Model No t Description Eff Disc y 1 2 3 4 5 6948 PROBE PACKAGE w termination 6949 PROBE PACKAGE w termination PROBE PACKAGE w termination 9 root probe probe package includes TERMINATION coaxial 6948 PROBE PACKAGE w accessories 6949 PROBE PACKAGE w accessories PROBE PACKAGE w accessories 9 foot probe probe package includes 6948 1 PROBE P6021 6949 1 PROBE P6021 1 PROBE P6201 9 foot 2 CLIP pro...

Page 18: ...on y 1 2 3 4 5 PROBE P6021 PROBE P6021 PROBE P6021 9 foot probe includes 1 BODY probe top 1 BALL metal 0 094 inch OD 1 SLIDE ASSEMBLY upper shell 1 CONTACT electrical 1 T1 TRANSFORMER 1 SPRING helical compression 1 RETAINER return spring 1 HOLDER spring return 2 SCREW 2 56 x 1 4 inch RHS 1 BODY probe bottom 1 BODY probe bottom 1 BODY probe bottom 9 foot probe 1 HOLDER circuit board 6 SOCKET pin co...

Page 19: ...TION Q Serial Model No Eff Disc Y 1 2 J 4 5 TERMINATION P6021 termination includes Description COVER termination top w knob CONNECTOR electrical BNC male FRAME termination CONNECTOR BNC female ASSEMBLY circuit board assembly includes INSULATOR plate BOARD circuit mounting hardware not included w assembly 2 SPACER sleeve 0 087 ID x 0 1 87 inch OD 2 WASHER fiat 0 090 ID x 0 188 inch OD 2 LOCKWASHER ...

Page 20: ... FORMER I I LI4 I 3T J R3b CoRE r 04 TI4 t P2 0 J 20 rl J R2ZJ n f f P39 LIO I I I I 1 _ 50M RIS 820 LI2 I I I I I I 5O u 1 4 CI3 2 I I I 1 ADDED AFTER 1 1 2 49 I C 50 C 2Z 5 25 L 31 t04 H R35 3 1 34 2 4 5 i III I I I II 5 ILLU L_____________L _____ _ I 1 I I I __ J 4 1 P o02 1 PROBE PG02 1 PA551VE TE RMINATIOt l I TYPE P60 2 J CURRENT PROBE TERMINATION 0670 R EV APR 1974 ...

Reviews: