Kit Description
Environmental Requirements
The following environmental characteristics supersede those listed in the
speci
fi
cations and performance veri
fi
cation sections of the manual for your
instrument.
Table 3: Warranted characteristics
Characteristic
Description
Temperature
inside rack cabinet
Operating
+10 °C to +45 °C (+50 °F to +113 °F)
Nonoperating
–22 °C to +60 °C (–7.6 °F to +140 °F)
Shock
Operating and
nonoperating
20 g, 11 ms, half sine, each axis, 3 drops each
Clearance Requirements
The rack in which the rack adapted instrument is mounted must provide the
following clearance requirements:
A minimum of 8.75 in (222.25 mm) of vertical space
A minimum width of 17.72 in (450.09 mm) between the left- and right-front
rails in the rack
A minimum depth of 12.75 in (323.85 mm)
CAUTION.
Adhering to these clearance requirements provides the rack-mounted
instrument with suf
fi
cient clearance for air circulation and accommodation of the
power cord and mounting hardware. Failure to provide these clearances can
result in overheating and can cause instrument faults or failure.
RMD5000 Rackmount Kit Instructions