In Case of Problems
TLA 700 Series Installation Manual
29
11. Refer to Figure 12 and bend the rubber comb to remove the faulty podlet
cable from the comb.
Remove faulty podlet cable from rubber comb
Cut above color-coded ring
Figure 12: Removing a podlet cable from the rubber comb
12. Attach the module end of the replacement podlet cable to the cut end of the
faulty podlet cable in the following manner:
a. Lay the two cables in parallel with a four-inch (10 cm) to six-inch
(15 cm) overlap.
b. Wrap the overlapping portion of the two podlet cables tightly with
masking tape to temporarily splice them together.
13. Pull the faulty cable from the module end through the mesh sleeve so that
the replacement cable is drawn through the mesh sleeve. Stop pulling when
the replacement cable is located at the proper position in the cable wire
bundle.
Summary of Contents for TLA 711
Page 4: ......
Page 8: ...Table of Contents iv TLA 700 Series Installation Manual...
Page 12: ...Service Safety viii TLA 700 Series Installation Manual...
Page 30: ...Basic Installation 14 TLA 700 Series Installation Manual...
Page 32: ...Installing Microprocessor Support Software 16 TLA 700 Series Installation Manual...
Page 34: ...The Next Step 18 TLA 700 Series Installation Manual...
Page 56: ...Appendix A Merging Logic Analyzer Modules 40 TLA 700 Series Installation Manual...