TEST Subsystem
2–68
TLA 720 Benchtop Chassis and TLA 7XM Expansion Chassis Service Manual
TEST?
This command executes all internal self tests. If a failure occurs and the halt
control is enabled, the test will immediately end. The query form returns the
unique numeric identifier of the first test that failed. A value of zero is returned if
there are no failures. The command form executes the same tests but returns no
results code.
Each internal self test has a unique numeric identifier in the range of 1000 to 1999
(see Table 2–11 on page 2–67).
*TST? and TEST? are aliases.
TEST[:ALL]
TEST[:ALL]?
Parameter
Query response
Not Applicable
<NR1>
0
1000
≤
N
≤
1999
Not Applicable
None
None
Command:
TEST
Query:
TEST?
Response:
0
TEST:RESults?
TEST:RESults:VERBose?
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands
Summary of Contents for TLA 720
Page 13: ...Table of Contents viii TLA 720 Benchtop Chassis and TLA 7XM Expansion Chassis Service Manual...
Page 22: ...Specifications...
Page 23: ......
Page 38: ...Operating Information...
Page 39: ......
Page 105: ...Command Groups 2 66 TLA 720 Benchtop Chassis and TLA 7XM Expansion Chassis Service Manual...
Page 111: ...TEST Subsystem 2 72 TLA 720 Benchtop Chassis and TLA 7XM Expansion Chassis Service Manual...
Page 124: ...Theory of Operation...
Page 125: ......
Page 128: ...Maintenance...
Page 129: ......
Page 155: ...Jumper Settings 4 26 TLA 720 Benchtop Chassis and TLA 7XM Expansion Chassis Service Manual...
Page 178: ...Options...
Page 179: ......
Page 182: ...Diagrams...
Page 183: ......
Page 186: ...Replaceable Mechanical Parts...
Page 187: ......
Page 198: ...Replaceable Parts TLA 720 Benchtop Chassis and TLA 7XM Expansion Chassis Service Manual 7 11...
Page 203: ...Replaceable Parts 7 16 TLA 720 Benchtop Chassis and TLA 7XM Expansion Chassis Service Manual...
Page 204: ...Index...
Page 205: ......
Page 211: ...Index Index 6 TLA 720 Benchtop Chassis and TLA 7XM Expansion Chassis Service Manual...
Page 212: ......
Page 213: ......