Mainframe Troubleshooting
4–42
TLA 720 Benchtop Chassis and TLA 7XM Expansion Chassis Service Manual
Isolating System Problems
If you have completed all of the troubleshooting procedures up to this point and
the TLA application fails to display any modules in the System Window, you
may have a system problem. Check for the following:
H
Verify that all modules are properly installed.
H
Verify that the module address switches are set correctly. Power off the
mainframe and remove the modules. Set the address switches to FF and
reinstall the modules.
H
Verify that the modules do not have the flash programming jumper installed
on the rear of the module. Power off the mainframe and remove the modules.
Remove the jumper and reinstall the modules.
H
Try placing a suspected module in a different slot to verify slot dependency
problems. For example, if you have a single module in slots 2 and 3, power
off the mainframe, move the module to slots 3 and 4, and try the tests again
If the module works in the new location, you have identified a faulty slot in
the mainframe.
H
Check for bent or broken pins on the backplane of the mainframe.
You can execute the internal resource manager program (ResMan32.exe) to
identify if any of the installed modules are being identified in the mainframe
slots. Table 4–6 lists some of the command line options for executing Res-
Man32.
Table 4–6: Command line options for ResMan32
Option
Description
–a, –A, –o ,–O
ResMan32 will not close the text window after executing and displaying
the results the major functions (default).
–p, –P
ResMan32 will not execute the mainframe power-on self test
diagnostics (default).
–v, –V
ResMan32 records the resource manager actions in the text window in
a short form or nonverbose mode.
+a, +A, +o ,+O
ResMan32 will terminate the tests and display the resultant action
information in the text window.
+p, +P
ResMan32 will perform the mainframe power-on self test diagnostics.
+v, +V
ResMan32 records all actions in a text window in the verbose mode
(default).
+t, +T
ResMan32 will not display the text window and the tests will terminate
after executing regardless of the error conditions.
+m, +M
ResMan32 displays in a minimized window.
Summary of Contents for TLA 720
Page 13: ...Table of Contents viii TLA 720 Benchtop Chassis and TLA 7XM Expansion Chassis Service Manual...
Page 22: ...Specifications...
Page 23: ......
Page 38: ...Operating Information...
Page 39: ......
Page 105: ...Command Groups 2 66 TLA 720 Benchtop Chassis and TLA 7XM Expansion Chassis Service Manual...
Page 111: ...TEST Subsystem 2 72 TLA 720 Benchtop Chassis and TLA 7XM Expansion Chassis Service Manual...
Page 124: ...Theory of Operation...
Page 125: ......
Page 128: ...Maintenance...
Page 129: ......
Page 155: ...Jumper Settings 4 26 TLA 720 Benchtop Chassis and TLA 7XM Expansion Chassis Service Manual...
Page 178: ...Options...
Page 179: ......
Page 182: ...Diagrams...
Page 183: ......
Page 186: ...Replaceable Mechanical Parts...
Page 187: ......
Page 198: ...Replaceable Parts TLA 720 Benchtop Chassis and TLA 7XM Expansion Chassis Service Manual 7 11...
Page 203: ...Replaceable Parts 7 16 TLA 720 Benchtop Chassis and TLA 7XM Expansion Chassis Service Manual...
Page 204: ...Index...
Page 205: ......
Page 211: ...Index Index 6 TLA 720 Benchtop Chassis and TLA 7XM Expansion Chassis Service Manual...
Page 212: ......
Page 213: ......