background image

Clear and Sanitize Procedures

To disable USB and CD-RW for Windows using the Windows Device Manager
complete the following steps:

1.

Connect a PS2 mouse and a PS2 keyboard to the instrument before powering
on (because USB will be disabled).

2.

Log on to the instrument as an administrator.

3.

From the Windows Start menu, go to

Control Panel

>

System

.

4.

Select the

Hardware

tab.

5.

Click

Device Manager

.

6.

Expand the Universal Serial Bus controllers entry by clicking the + next to it.

7.

Double-click the

fi

rst USB Root Hub entry.

8.

Select the Power tab.

9.

If the Device Description is anything other than Generic USB Hub (4 ports),
click the General tab, and select Do not use this device (disable) in the Device
Usage drop-down list.

NOTE.

It is critical to leave the USB Root Hub operating, otherwise, the front

panel will not function. If the front panel does not function, the TLA application
will not run.

10.

Click OK.

11.

Repeat steps through 9 for each USB Root Hub shown in the Device
Manager window.

12.

Expand DVD/CD-ROM drives in the Device Manager window.

13.

Right-click the drive, and select Disable.

14.

Restart the instrument to make the changes effective.

15.

Tektronix suggests that you password-protect the Windows Administrator
account and set up Guest accounts for end users so that these changes cannot
be easily reversed.

To Disable LAN Ethernet

Connectivity Using the

BIOS

1.

Press Delete during instrument Boot Up to go to the BIOS con

fi

guration menu.

2.

Go to

Advanced

>

Peripheral Con

fi

guration.

3.

Set Onboard LAN to Disabled.

4

TLA6000 Declassi

fi

cation and Security Instructions

Summary of Contents for TLA6000 Series

Page 1: ...TLA6000 Series Logic Analyzers Declassification and Security ZZZ Instructions xx www tektronix com P077048300 077 0483 00...

Page 2: ...tion in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of Tektronix Inc TekLink i...

Page 3: ...Memory Devices 1 Data Export Devices 3 Built In Security Features 6 Troubleshooting 7 How to Clear or Sanitize a Nonfunctional Instrument 7 How to Recover from Clearing or Removing the Instrument s M...

Page 4: ...Table of Contents ii TLA6000 Declassification and Security Instructions...

Page 5: ...l TLA7ACx Logic Analyzer Module Service Manual Terms The following terms may be used in this document Clear This removes data on media memory before reusing it in a secured area All reusable memory is...

Page 6: ...temporarily or permanently from a secured area to a non secured area Scrub This is equivalent to sanitize User modifiable The memory device can be written to by the user during normal instrument opera...

Page 7: ...storing acquired data No Written by controller ASICs Acquisition board Remove power source from the instrument for at least 20 seconds Remove power source from the instrument for at least 20 seconds S...

Page 8: ...therboard Load new BIOS from vendor Web site Securely store instrument in a safe location or destroy it CMOS Configurable BIOS parameters No Written by the processor system Motherboard Remove battery...

Page 9: ...d rear of instrument Files can be deleted or overwritten on the instrument or a PC USB flash drive can be removed and destroyed The USB ports can be disabled See page 3 Disabling USB and CD RW Capabil...

Page 10: ...e Device Usage drop down list NOTE It is critical to leave the USB Root Hub operating otherwise the front panel will not function If the front panel does not function the TLA application will not run...

Page 11: ...ck the General tab and select Do not use this device disable in the Device Usage drop down list Removable Hard Disk Drive To Remove the Hard Disk Drive To remove the hard disk drive assembly refer to...

Page 12: ...f the hard disk drive has been removed and a new hard disk drive is installed you need to install the operating system and the TLA application software Refer to the TLA6000 Series Logic Analyzer Insta...

Page 13: ...tached USB flash drive from the instrument before returning the instrument for repair If your instrument is not functioning and you need to clear it remove the power source from the instrument for at...

Reviews: