Mainframe Removal and Installation Procedures
TLA721 Benchtop & TLA7XM Expansion Mainframe Service Manual
6- 19
EMI DIN Shields
Perform the following steps to remove the EMI DIN shields:
1.
Remove any modules surrounding the slots where you intend to remove the
Backplane EMI DIN shields. Refer to Figure 6--11.
2.
Remove two 4-40 Torxdrive T-10 screws that secure each EMI DIN shield to
the backplane.
3.
Remove EMI DIN shield from the backplane.
Install the EMI DIN shield by reversing the removal procedure.
Screws (2)
Polarization hole
Backplane EMI
DIN shield
Figure 6- 11: Removing the backplane EMI DIN shields
Removal
Installation
Summary of Contents for TLA7XM
Page 5: ......
Page 13: ...Table of Contents viii TLA721 Benchtop TLA7XM Expansion Mainframe Service Manual ...
Page 17: ...Service Safety Summary xii TLA721 Benchtop TLA7XM Expansion Mainframe Service Manual ...
Page 21: ...Preface xvi TLA721 Benchtop TLA7XM Expansion Mainframe Service Manual ...
Page 25: ...Introduction xx TLA721 Benchtop TLA7XM Expansion Mainframe Service Manual ...
Page 43: ...Specifications 1 18 TLA721 Benchtop TLA7XM Expansion Mainframe Service Manual ...
Page 85: ......
Page 87: ...Adjustment Procedures 5 2 TLA721 Benchtop TLA7XM Expansion Mainframe Service Manual ...
Page 161: ...Repackaging Instructions 6 74 TLA721 Benchtop TLA7XM Expansion Mainframe Service Manual ...
Page 165: ...Electrical Parts List 8 2 TLA721 Benchtop TLA7XM Expansion Mainframe Service Manual ...
Page 169: ...Diagrams 9 4 TLA721 Benchtop TLA7XM Expansion Mainframe Service Manual ...
Page 189: ...Mechanical Parts List 10 20 TLA721 Benchtop TLA7XM Expansion Mainframe Service Manual ...