Specifications
1- 6
TPS2000 Series Digital Storage Oscilloscopes Service Manual
Table 1- 1: Oscilloscope general characteristics (Cont.)
Pulse Width Trigger
Pulse Width Trigger modes
Trigger when < (Less than), > (Greater than), = (Equal), or
≠
(Not Equal);
Positive pulse or Negative pulse.
Pulse Width Trigger Point
Equal:
The oscilloscope triggers when the trailing edge of the pulse crosses the trigger level.
Not Equal:
If the pulse is narrower than the specified width, the trigger point is the trailing
edge. Otherwise, the oscilloscope triggers when a pulse continues longer than the time
specified as the Pulse Width.
Less than:
The trigger point is the trailing edge.
Greater than
(also called time-out trigger): The oscilloscope triggers when a pulse continues
longer than the time specified as the Pulse Width.
Pulse Width Range
Selectable from 33 ns to 10 s
Pulse Width Resolution
16.5 ns or 1 part per thousand, whichever is larger
Equal Guard Band
t > 330 ns:
±
5%
≤
guard band <
±
(5.1% + 16.5 ns)
t
≤
330 ns: guard band =
±
16.5 ns
Not Equal Guard Band
t > 330 ns:
±
5%
≤
guard band <
±
(5.1% + 16.5 ns)
165 ns < t
≤
330 ns: guard band = --16.5 ns/+33 ns
t
≤
165 ns: guard band =
±
16.5 ns
Trigger Frequency Counter
Readout Resolution
6 digits
Accuracy, typical
±
51 ppm including all frequency reference errors and
±
1 count errors
Frequency Range
AC coupled, 10 Hz minimum to rated bandwidth.
Signal Source
Pulse Width/ Edge Trigger:
all available trigger sources.
Frequency Counter
measures trigger source at all times, including when the oscilloscope
acquisition is halted due to changes in the run status, or acquisition of a single shot event
has completed.
Pulse Width Trigger:
The oscilloscope counts pulses of significant magnitude inside the 250
ms measurement window that qualify as triggerable events, such as narrow pulses in a PWM
pulse train if set to < mode and the width is set to a relatively small time.
Edge Trigger:
The oscilloscope counts all edges of sufficient magnitude and correct polarity.
Video Trigger:
The Frequency Counter does not operate.
Summary of Contents for TPS2000 Series
Page 6: ......
Page 12: ...Table of Contents vi TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 16: ...Service Safety Summary x TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 20: ...Preface xiv TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 42: ...Theory of Operation 3 6 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 62: ...Adjustment Procedures 5 10 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 90: ...Maintenance 6 28 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 92: ...Options 7 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 94: ...Electrical Parts List 8 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 96: ...Diagrams 9 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 106: ...Replaceable Parts 10 10 TPS2000 Series Digital Storage Oscilloscopes Service Manual...