To remove the IO board, refer to Figure 6--2 on page 6--10 and follow these steps:
1.
Remove the rear case. Follow the procedure on page
6--7
.
2.
Disconnect all cables from the IO board.
3.
Remove the three chassis mounting screws from the IO board.
4.
Unseat the IO board from its connector on the acquisition board and lift up to
remove the board.
Installation.
To install a new IO board, refer to Figure 6--2 and follow these steps:
1.
Align the connector on the bottom of the IO board with the connector on the
acquisition board. Press down to seat the connector.
2.
Install the three IO board mounting screws to attach the board to the support
bracket.
3.
Connect all cables to the IO board.
4.
Install the rear case.
Cables
Backlight
Inverter board
Fan
IO board
Acquisition board
Display data cable
I/O board screws (3)
Figure 6--2: Modules accessed from the rear of the oscilloscope
IO Board
Summary of Contents for TPS2000 Series
Page 6: ......
Page 12: ...Table of Contents vi TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 16: ...Service Safety Summary x TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 20: ...Preface xiv TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 42: ...Theory of Operation 3 6 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 62: ...Adjustment Procedures 5 10 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 90: ...Maintenance 6 28 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 92: ...Options 7 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 94: ...Electrical Parts List 8 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 96: ...Diagrams 9 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 106: ...Replaceable Parts 10 10 TPS2000 Series Digital Storage Oscilloscopes Service Manual...