VX1411 MainFrame Instruction Manual
4–1
Specifications
This chapter contains the complete specifications for the VX1411 Mainframe.
Within each section, the specifications are arranged in functional groups such as:
AC Power Source, Secondary Power, Cooling, Safety, Environmental, Back-
plane, and Mechanical.
All specifications are warranted unless they are designated typical. DC Voltage
Regulation is a warranted characteristic and can be directly checked by proce-
dures contained in the Performance Verification chapter beginning on page 5–1;
DC Voltage Regulation is listed in boldface under the Characteristics column.
If the characteristic is noted as typical, the characteristic is not warranted.
Typical characteristics describe typical or average performance and provide
useful reference information.
Table 4–1: AC power source
Characteristic
Description
Source Voltage
90 V
RMS
to 250 V
RMS
, continuous range
Source Frequency
45 Hz to 66 Hz or 360 Hz to 440 Hz, 115 V operation
45 Hz to 66 Hz, 230 V operation
Power Consumption
800 W line power maximum
Fuse Rating
90 V – 250 V Operation
0.25 in
×
1.25 in, Fast Blow, 10 A, 250 V
207 V – 250 V Operation
5 mm
×
20 mm, Slow Blow, 4 A, 250 V
Inrush Surge Current
70 A maximum
Input Current
8 A maximum at 90 V
RMS
3.5 A maximum at 207 V
RMS
Power Factor Correction
Typically 0.99 at 60 Hz operation and 0.95 at 400 Hz
operation
Power Disconnect
Front Panel On/Standby. No primary switch on rear
panel. Power cord provides main power disconnect.