Specifications
4–2
VX1411 MainFrame Instruction Manual
Table 4–2: Secondary power
Characteristic
Description
Maximum Power Available
Mainframe (Power available to
modules)
500 W, 0
°
C to 55
°
C
DC Current Capacity (I
MP
)
Voltage
I
MP
(steady-state current)
+24 V
7 A
+12 V
6 A
+5 V
60 A
–2 V
–10 A
–5.2 V
–20 A
–12 V
–6 A
–24 V
–7 A
AC Current Capacity (I
MD
)
Voltage
I
MD
(dynamic current)
+24 V
5 A
+12 V
2 A
+5 V
10 A
–2 V
–5 A
–5.2 V
–7 A
–12 V
–2 A
–24 V
–5 A
DC voltage regulation
Voltage
Tolerance,
V
+24 V
+1.2 V, –720 mV
+12 V
+600 mV, –360 mV
+5 V
+250 mV, –125 mV
–2 V
–100 mV, +100 mV
–5.2 V
–260 mV, +156 mV
–12 V
–600 mV, +360 mV
–24 V
–1.2 V, +720 mV