Portable Trace Oxygen Analyzer
Operation
Teledyne Analytical Instruments
15
Operation
The Model 3110XL is supplied completely assembled and ready
for use. The micro-fuel cell is supplied in a separate sealed bag, and in
the case of trace sensors, was purged with an inert gas.
Turning the instrument on by pressing the ENTER key will power
the display and show the power on screen briefly (if the display shows
nothing or is very dark, adjust the contrast). The display will then
change to indicate the oxygen concentration within the internal sample
passageways if no sample line is attached.
Note: To extend cell life and minimize the time required to make
the next analysis, the instrument should always be purged
with an oxygen free inert gas prior to being taken out of
service for standby or storage.
3.1 Front Panel Interface
The Model 3110XL is controlled from the keys on the front panel
and is shown in Figure 3-1. These keys are also used to setup the
instrument for your application. The keys are:
•
ENTER/ON
•
ESC/OFF
•
UP
•
DOWN
3.1.1 ENTER Key
The ENTER key is context sensitive. It is used as follows:
•
Powering ON or OFF
—Pressing the ENTER key turns the
power ON. The ENTER key is also used the power OFF
from within the
POWER OFF
screen.
•
Enter SETUP
—In certain menus, pressing the ENTER key
selects a setup screen for that particular function. Navigation
arrows on the left and right of the bottom line identify these
screens as setup screens by changing from UP/DOWN to
LEFT/RIGHT. The setup screens also blink.