Operation
Teledyne API M751H High Performance Zero Air Generator
26
07773B DCN6940
Provision for voltage configuration is via a jumper plug (J5). All electrical connections
are made by quick release connectors to aid in servicing. Refer to Figure 4-4 for the
layout and to Appendix B for the interconnect and other drawings.
The 751H Control PCB is microcontroller based and provides the following functions:
•
Cycles the four-way valve of the Regenerative Dryer.
•
Cycles the water drain solenoid valve.
•
Cycles the relief valve in response to the pressure sensor.
•
Cycles the Regenerative Dryer four-way valve and momentarily opens the water
drain solenoid valve at compressor startup. This reduces the pressure at the
compressor outlet to facilitate compressor starting.
LED D13 flashes at 1 second intervals as an indication that the controller is
functioning, and a watchdog timer is enabled to prevent any power line disturbances
from halting the processor.
Figure 4-4: Control Board Layout
Summary of Contents for 751H
Page 2: ......
Page 4: ...ii 07773B DCN6940 This page intentionally left blank...
Page 44: ...A 2 07773B DCN6940 This page intentionally left blank...
Page 46: ...This page intentionally left blank A 4 07773B DCN6940...
Page 47: ...07773B DCN6940 B 1 APPENDIX B SCHEMATICS...
Page 48: ...B 2 07773B DCN6940 This page intentionally left blank...