MAINTENANCE
Teledyne API M751H High Performance Zero Air Generator
38
07773B DCN6940
5.12 Filter Element Replacement
5.12.1 Intake Filter
To service the intake filter (Figure 4-2):
1. Turn the filter cap counter-clockwise and remove from the filter housing.
2. Remove the filter screen and use compressed air to blow out any debris.
3. Reseat the filter screen inside the filter housing.
4. Return the cap to the filter housing and secure in place with a clockwise turn.
If the screen is damaged, then replace the intake filter assembly (TAPI P/N FL51).
5.12.2 Final Filter
To replace the disposable Final filter unit (Figure 4-2):
1. Remove the disposable filter unit (DFU) from the retainer clip.
2. Release the tubing at each end of the DFU by pressing the spring-loaded quick-
connects toward the DFU and pulling the tubing away.
3. Dispose of the used DFU.
4. Reinsert the tubing into each end of the new DFU (Teledyne API P/N FL50),
pushing in tightly to secure.
5. Secure the new filter unit into the retainer clip.
Summary of Contents for 751H
Page 2: ......
Page 4: ...ii 07773B DCN6940 This page intentionally left blank...
Page 44: ...A 2 07773B DCN6940 This page intentionally left blank...
Page 46: ...This page intentionally left blank A 4 07773B DCN6940...
Page 47: ...07773B DCN6940 B 1 APPENDIX B SCHEMATICS...
Page 48: ...B 2 07773B DCN6940 This page intentionally left blank...