10.
In
preparation
for installing the new thermocouple, look into the end of the Heater Block. You will
see that there is an indentation (cavity) in the fat part of the quartz tube. This is where the
thermocouple you are installing will reside. Refer to Figure 2-3.
Figure 2-3. Quartz Tube Cavity for Thermocouple
11.
The
thermocouple should slide into the Heater Block and into the indentation of the quartz.
12.
Align the thermocouple with this cavity and carefully push the thermocouple all the way into the
cavity until it comes to a stop, which is the end of the cavity of the quartz tube.
13.
The thermocouple should now be properly seated in the cavity of the quartz tube. Refer to Figure
Figure 2-4. Thermocouple Installed
14.
Reconnect
the Teflon fitting that was removed earlier from the end of the quartz tube. Take care
not to put any stress on the quartz tube as the Teflon fitting is tightened.