Setup Menu
Teledyne API – Model T300/T300M CO Analyzer
108
5.9.1.
SIGNAL I/O
The signal I/O diagnostic mode allows a user to review and change the digital and
analog input/output functions of the analyzer. Refer to Appendix A-4 for a full list of
the parameters available for review under this menu.
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
Any changes of signal I/O settings will remain in effect only until the
signal I/O menu is exited. Exceptions are the ozone generator override
and the flow sensor calibration, which remain as entered when exiting.
Access the SIGNAL I/O test mode from the DIAG Menu:
SAMPLE
RANGE=50.0 PPM CO= XXXX
<TST
TST> CAL
SETUP
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
DIAG
EXIT
EXAMPLE
DIAG I/O 0) EXT_ZERO_CAL=OFF
PREV
NEXT
EDIT PRNT
EXIT
Press
PREV
and
NEXT
to cycle through
the signal types.
DIAG I/O 1)EXT_SPAN_CAL=OFF
PREV NEXT JUMP
EDIT PRNT
EXIT
Press
JUMP
to go
directly to a specific
signal
(see Appendix A for a list
of all
I/O SIGNALS
)
DIAG I/O JUMPTO: 0
0
0
JUMP
ENTR
EXIT
Toggle to set
location No. of the
signal to
JUMP
to.
DIAG I/O JUMPTO: 22
2
2
JUMP
ENTR
EXIT
DIAG I/O 22) RELAY_WATCHDOG=ON
PREV NEXT JUMP
OFF PRNT EXIT
Pressing
PRNT
will send a formatted
printout to the serial port and can be
captured with a computer or other
output device.
SETUP X.X
ENTER PASSWORD:818
8
1
8
ENTR
EXIT
DIAG SIGNAL I/O
NEXT
ENTR
EXIT
On status signals this
button toggles the
signal
ON / OFF.
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
06864D DCN7562