Teledyne API – Model T300/T300M CO Analyzer
Setup Menu
125
5.9.3.10.
AIN CALIBRATION
This is the submenu to conduct a calibration of the T300/T300M Analyzer’s analog
inputs. This calibration should only be necessary after major repair such as a
replacement of CPU, motherboard or power supplies.
To perform an analog input calibration, select the
ANALOG I/O CONFIGURATION
submenu (see Figure 5-3) then press:
DIAG AIO AIN CALIBRATED:NO
<SET
CAL
EXIT
DIAG
ANALOG I/O CONFIGURATION
PREV NEXT
ENTR
EXIT
Firmware automatically performs a zero point calibration
of the Motherboard’s analog Inputs
Continue pressing SET> until you reach the
output to be configured
DIAG AIO AOUTS CALIBRATED: NO
<SET
SET>
CAL
EXIT
Firmware automatically performs a span point calibration
of the Motherboard’s analog Inputs
DIAG AIO CALIBRATING A/D ZERO
DIAG AIO CALIBRATING A/D SPAN
DIAG AIO A/D CALIBRATION ERROR
EXIT
DIAG AIO AIN CALIBRATED: YES
EXIT
Perform
Troubleshooting or call
Teledyne Customer
Service.
From the
AIO CONFIGURATION SUBMENU
06864D DCN7562