Teledyne API – Model T300/T300M CO Analyzer
Data Acquisition System (DAS) and APICOM
175
7.1.11.
HOLDOFF FEATURE
The DAS
HOLDOFF
feature prevents data collection during calibration operations and
at certain times when the quality of the analyzer’s CO measurements may not be certain
(e.g. while the instrument is warming up). In this case, the length of time that the
HOLDOFF
feature is active is determined by the value of the internal variable (
VARS
),
DAS_HOLDOFF.
To set the length of the DAS_HOLDOFF period, go to the SETUP>MORE>VARS
menu (Section 5.8) and EDIT the “0) DAS_HOLD_OFF…” parameter.
To enable or disable the
HOLDOFF
feature for an individual channel, follow the
instruction shown in Section 7.1.4 then press:
Starting at the
EDIT CHANNEL MENU
SETUP X.X
0) CONC: ATIMER 1, 800
PREV MEXT
INS
DEL
EDIT
PRNT
EXIT
SETUP X.X
NAME: CONC
<SET
SET>
EDIT
EXIT
Continue pressing
SET>
until ...
SETUP X.X
CAL.HOLD OFF: OFF
<SET
SET
>
EDIT
EXIT
Use the
PREV
and
NEXT
buttons to
scroll to the
DATA
CHANNEL
to be
edited.
SETUP X.X
CAL.HOLD OFF: OFF
OFF
ENTR EXIT
EXIT
discards the new
setting.
ENTR
accepts the
new setting.
Toggle these buttons
to turn the
HOLDOFF
feature
ON/OFF.
06864D DCN7562