Calibration Procedures
Teledyne API – Model T300/T300M CO Analyzer
202
SAMPLE RANGE=50.0PPM CO=XX.XX
< TST TST > CAL CALZ
CALS
SETUP
Only appears if either
the O
2
or or the CO
2
sensor is installed.
M-P CAL CO SPAN CONC:40.0 Conc
0
0
4
0
.0
0
ENTR EXIT
SAMPLE
GAS TO CAL:CO
CO
O2
ENTR
EXIT
M-P CAL RANGE=50.0PPM CO=XX.XX
<TST TST> ZERO SPAN
CONC
EXIT
SAMPLE
RANGE TO CAL:LOW
LOW
HIGH
ENTR
EXIT
The default CO span concentration
value is 4
0.0 Conc
.
If this is not the the concentration of
the span gas being used, toggle
these buttons to set the correct
concentration of the CO
calibration gas.
EXIT
ignores the new
setting and returns to
the previous display.
ENTR
accepts the new
setting and returns to
the
CONCENTRATION
MENU.
Only appears if the
DUAL
or
AUTO
range modes are selected.
Use these buttons to choose the
appropriate range.
Repeat entire procedure for each
range.
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
For this Initial Calibration it is important to independently verify the
PRECISE CO Concentration Value of the SPAN gas.
If the source of the Span Gas is from a Calibrated Bottle, use the exact
concentration value printed on the bottle.
9.3.3.2.
ZERO/SPAN POINT CALIBRATION PROCEDURE
The zero and cal operations are initiated directly and independently with dedicated
buttons (
CALZ
&
CALS
).
06864D DCN7562