Calibration Procedures
Teledyne API – Model T300/T300M CO Analyzer
214
To cause the analyzer to measure and record a value for
PRES
, press.
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
SETUP X.X
ENTER PASSWORD:818
8
1
8
ENTR
EXIT
DIAG SIGNAL I/O
NEXT
ENTR
EXIT
Continue pressing NEXT until ...
EXIT discards the
new setting.
ENTR accepts the
new setting.
DIAG
PRESSURE CALIBRATION
PREV NEXT
ENTR
EXIT
SAMPLE
RANGE=50.0 PPM CO= XX.XX
<TST TST> CAL
SETUP
DIAG PCAL
PRESSURE CALIBRATION
CAL
EDIT
EXIT
DIAG PCAL
SAMPLE PRESS: 29.93 IN-HG-A
2
9
.9
3
ENTR EXIT
06864D DCN7562