Teledyne API – Model T300/T300M CO Analyzer
Calibration Procedures
215
9.6.3.
FLOW CALIBRATION
The flow calibration allows the user to adjust the values of the sample flow rates as they
are displayed on the front panel and reported through COMM ports to match the actual
flow rate measured at the sample inlet. This does not change the hardware measurement
of the flow sensors, only the software-calculated values.
To carry out this adjustment, connect an external, sufficiently accurate flow meter to the
sample inlet (see Section 10.3.4 for more details). Once the flow meter is attached and
is measuring actual gas flow, press:
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
DIAG
EXIT
SETUP X.X
ENTER PASSWORD:818
8
1
8
ENTR
EXIT
DIAG SIGNAL I/O
PREV NEXT
ENTR
EXIT
Continue pressing
NEXT
until ...
DIAG FCAL
ACTUAL FLOW: 1000 CC/M
1
0
0
0
ENTR
EXIT
Toggle these buttons to
match the actual flow as
measured by the external
flow meter.
EXIT
discards the new
setting.
ENTR
accepts the
new setting.
DIAG
FLOW CALIBRATION
PREV NEXT
ENTR
EXIT
SAMPLE
RANGE=50.0 PPM CO= XX.XX
<TST
TST> CAL
SETUP