Teledyne API – Model T300/T300M CO Analyzer
Calibration Procedures
221
9.7.2.2.
SET CO2 SPAN GAS CONCENTRATION:
Set the expected CO
2
span gas concentration.
This should be equal to the percent concentration of the CO
2
span gas of the selected
reporting range (default factory setting = 12%).
SAMPLE
GAS TO CAL:CO
CO
CO2
O2
ENTR EXIT
SAMPLE RANGE=50.0 PPM CO= XX.XX
< TST TST >
CAL
SETUP
M-P CAL RANGE=50.0 PPM CO= XX.XX
<TST TST> ZERO SPAN
CONC
EXIT
SAMPLE
GAS TO CAL:O2
CO
CO2
O2
ENTR
EXIT
M-P CAL C O2 SPAN CONC:12.00%
0
1
2
.0
0
ENTR EXIT
The CO
2
span concentration value automatically defaults
to 12 %.
If this is not the the concentration of the span gas being
used, toggle these buttons to set the correct concentration
of the CO
2
calibration gases.
EXIT ignores the new
setting and returns to
the previous display.
ENTR accepts the new
setting and returns to
the previous menu.
06864D DCN7562