Troubleshooting and Service
Teledyne API – Model T300/T300M CO Analyzer
238
Suppresses the
warning messages.
Press
CLR
to clear the current
message.
If more than one warning is
active, the next message will take
its place.
Once the last warning has
been cleared, the
RANGE
function will be displayed in
the analyzer’s main
MESSAGE FIELD.
NOTE:
If a warning message persists after
several attempts to clear it, the message
may indicate a real problem and not an
artifact of the warm-up period.
SAMPLE SYSTEM RESET
TEST
CAL
MSG CLR SETUP
SAMPLE SYSTEM RESET
TEST
CAL
MSG CLR SETUP
SAMPLE SYSTEM RESET
TEST
CAL
MSG CLR SETUP
SYSTEM SYSTEM RESET
TEST
CLR
SETUP
MSG
returns the active
warnings to the message
field.
STANDBY
RANGE=50.0 PPB CO=XX.XX
TEST
CAL
MSG
SETUP
Figure 11-1:
Viewing and Clearing Warning Messages
06864D DCN7562