Teledyne API – Model T300/T300M CO Analyzer
Specifications and Approvals
25
Table 2-3:
CO
2
Sensor Option Specifications
Parameter
Description
Ranges
0-1% to 0-20% user selectable. Dual ranges and auto-
ranging supported.
Zero Noise
1
<0.02% CO
2
Lower Detectable Limit
2
<0.04% CO
2
Zero Drift (24 hours)
<± 0.02% CO
2
Zero Drift (7 days)
<± 0.05% CO
2
Span Noise
1
<± 0.1% CO
2
Span Drift (7 days)
<± 0.1% CO
2
Accuracy
<± (1.5% of range + 2% of reading)
Linearity
<± 0.1% CO
2
Temperature Coefficient
<± 0.01% CO
2
/°C
Rise and Fall Time
<60 seconds to 95%
1
As defined by the USEPA
2
Defined as twice the zero noise level by the USEPA
2.2.
EPA DESIGNATION
Note
T300M: EPA designation does not apply to this model.
Teledyne API’s Model T300, Gas Filter Correlation Carbon Monoxide Analyzer, is
designated as a US EPA Reference Method, Designation Number RFCA-1093-093 for
CO measurement. The official “List of Designated Reference and Equivalent Methods”
is published in the U.S. Federal Register:
http://www3.epa.gov/ttn/amtic/criteria.html
2.3.
APPROVALS AND CERTIFICATIONS
The Teledyne API Model T300/T300M analyzer was tested and certified for Safety and
Electromagnetic Compatibility (EMC). This section presents the compliance statements
for those requirements and directives.
2.3.1.
SAFETY
IEC/EN 61010-1:2010 (3
rd
Edition), Safety requirements for electrical equipment for
measurement, control, and laboratory use.
CE:
2006/95/EC, Low-Voltage Directive
2.3.2.
EMC
EN 61326-1 (IEC 61326-1), Class A Emissions/Industrial Immunity
EN 55011 (CISPR 11), Group 1, Class A Emissions
FCC 47 CFR Part 15B, Class A Emissions
CE:
2004/108/EC, Electromagnetic Compatibility Directive
06864D DCN7562
Summary of Contents for T300
Page 2: ......
Page 182: ...06864D DCN7562 ...
Page 227: ...225 This page intentionally left blank 06864D DCN7562 ...
Page 228: ...06864D DCN7562 ...