Teledyne API - T300/T300M and M300E/EM 04906J DCN7122
APPENDIX A-2: Setup Variables For Serial I/O
A-9
APPENDIX A-2: Setup Variables For Serial I/O
Table A-1: T300/T300M and M300E/EM Setup Variables, Revision L.8
Setup Variable
Numeric
Units
Default
Value
Value
Range
Description
DAS_HOLD_OFF
Minutes
15
0.5–20
Duration of DAS hold off period.
CONC_PRECISION
—
3
AUTO,
0,
1,
2,
3,
4
Number of digits to display to the
right of the decimal point for
concentrations on the display.
REM_CAL_DURATION
1
Minutes
20
1–120
Duration of automatic calibration
initiated from TAI protocol.
STABIL_GAS
—
CO
0
CO,
CO2
2
O2
3
Selects gas for stability
measurement.
DYN_ZERO
—
OFF
ON, OFF
ON
enables remote dynamic
zero calibration;
OFF
disables it.
DYN_SPAN
—
OFF
ON, OFF
ON
enables remote dynamic
span calibration;
OFF
disables it.
CLOCK_ADJ
Sec./Day
0
-60–60
Time-of-day clock speed
adjustment.
SERVICE_CLEAR
OFF
OFF
ON
ON resets the service interval
timer.
TIME_SINCE_SVC
Hours
0
0-500000
Time since last service.
SVC_INTERVAL
Hours
0
0-10000
Sets the interval between service
reminders.
0
Enclose value in double quotes (") when setting from the RS-232 interface
1
TAI protocol
2
CO
2
option
3
O
2
option
06864D DCN7562
Summary of Contents for T300