APPENDIX A-7: MODBUS Register Map
Teledyne API - T300/T300M and M300E/EM 04906J DCN7122
A-26
MODBUS
Register Address
(dec., 0-based)
Description
Units
MODBUS Coil Registers
(single-bit; read/write)
0
Maps to relay output signal 36 (
MB_RELAY_36
in signal I/O list)
1
Maps to relay output signal 37 (
MB_RELAY_37
in signal I/O list)
2
Maps to relay output signal 38 (
MB_RELAY_38
in signal I/O list)
3
Maps to relay output signal 39 (
MB_RELAY_39
in signal I/O list)
20
4
Triggers zero calibration of range #1 (on enters cal.; off exits cal.)
21
4
Triggers span calibration of range #1 (on enters cal.; off exits cal.)
22
4
Triggers zero calibration of range #2 (on enters cal.; off exits cal.)
23
4
Triggers span calibration of range #2 (on enters cal.; off exits cal.)
1
Optional
2
O
2
option
3
CO
2
option
4
Set
DYN_ZERO
or
DYN_SPAN
variables to
ON
to enable calculating new slope or offset. Otherwise a calibration check
is performed.