Teledyne API – Model T300/T300M CO Analyzer
Getting Started
41
Figure 3-10:
Current Loop Option Installed on Motherboard
C
ONVERTING
C
URRENT
L
OOP
A
NALOG
O
UTPUTS TO
S
TANDARD
V
OLTAGE
O
UTPUTS
To convert an output configured for current loop operation to the standard 0 to 5 VDC
output operation:
1. Turn off power to the analyzer.
2. If a recording device was connected to the output being modified, disconnect it.
3. Remove the top cover.
•
Remove the screw located in the top, center of the front panel.
CAUTION – AVOID INVALIDATING WARRANTY
Servicing or handling of circuit components requires electrostatic
discharge protection, i.e. ESD grounding straps, mats and containers.
Failure to use ESD protection when working with electronic assemblies will
void the instrument warranty. Refer to the manual, Fundamentals of ESD,
PN 04786, which can be downloaded from our website at
under Help Center > Product Manuals in the
Special Manuals section.