Teledyne API – Model T300/T300M CO Analyzer
Getting Started
49
Figure 3-15:
Default Pin Assignments for CPU COM Port connector (RS-232)
RS-232
C
OM
P
ORT
D
EFAULT
S
ETTINGS
Received from the factory, the analyzer is set up to emulate a DCE or modem, with Pin
3 of the DB-9 connector designated for receiving data and Pin 2 designated for sending
data.
•
RS-232 (COM1)
: RS-232 (fixed) DB-9 male connector.
•
Baud rate
: 115200 bits per second (baud)
•
Data Bits
: 8 data bits with 1 stop bit
•
Parity
: None
•
COM2
: RS-232 (configurable to RS-485), DB-9 female connector.
•
Baud rate
: 19200 bits per second (baud)
•
Data Bits
: 8 data bits with 1 stop bit
•
Parity
: None
RS-232
M
ULTIDROP
O
PTION
C
ONNECTION
When the RS-232 Multidrop option is installed, connection adjustments and
configuration through the menu system are required. This section provides instructions
for the internal connection adjustments, then for external connections, and ends with
instructions for menu-driven configuration.
06864D DCN7562
Summary of Contents for T300