Teledyne API – Model T300/T300M CO Analyzer
Getting Started
63
E
XHAUST
O
UTLET
Attach an exhaust line to the analyzer’s EXHAUST outlet fitting. The exhaust line
should be:
•
PTEF tubing; minimum O.D ¼”;
•
A maximum of 10 meters long;
•
Vented outside the analyzer’s enclosure.
3.3.2.8.
PNEUMATIC LAYOUT FOR ZERO SCRUBBER/PRESSURIZED SPAN OPTION
INSTRUMENT CHASSIS
Flow / Pressure
Sensor PCA
SAMPLE
PRESSURE
SENSOR
FLOW
SENSOR
Sample Gas
Flow Control
SAMPLE
GAS INLET
EXHAUST
GAS OUTLET
Particulate
Filter
PUMP
ZERO AIR
INLET
PRESSURE
SPAN INLET
VENT SPAN
OUTLET
Shutoff
Valve
Sample / Cal
Valve
Zero / Span
Valve
GFC Wheel
Housing
GFC Motor Heat
Sync
SAMPLE CHAMBER
ZERO Air
Scrubber
Ambient
Sample
Gas
In
Span
Gas
In
Figure 3-26:
Internal Pneumatic Flow OPT 50E – Zero Scrubber/Pressurized Span
Table 3-10: Zero/Span Valve Operating States for Option 51E
Mode
Valve
Condition
SAMPLE
(Normal State)
Sample/Cal
Open to SAMPLE inlet
Zero/Span
Open to internal ZERO AIR
scrubber
Shutoff Valve
Closed
ZERO CAL
Sample/Cal
Open to zero/span valve
Zero/Span
Open to internal ZERO AIR
scrubber
Shutoff Valve
Closed
SPAN CAL
Sample/Cal
Open to ZERO/SPAN valve
Zero/Span
Open to SHUTOFF valve
Shutoff Valve
Open to PRESSURE SPAN
inlet
06864D DCN7562