Teledyne API – Model T300/T300M CO Analyzer
Overview of Operating Modes
81
1
This will match the currently selected units of measure for
the range being displayed.
2
Only appears if CO
2
sensor option is installed.
3
Only appears if O
2
sensor option is installed.
4
The
STABIL
function can be set to display data related to
any of the gasses the analyzer measures, e.g. (if either
the CO
2
or the O
2
sensor options are installed.
5
Only appears if analog output
A4
is actively reporting a
TEST FUNCTION
SAMPLE
RANGE=50.0 PPM CO= XX.XX
<TST
TST> CAL
SETUP
Toggle
<TST TST>
to scroll through list of
functions.
•
RANGE=[Value]PPB
1
•
RANGE1=[Value]PPB
1
•
RANGE2=[Value]PPB
1
•
CO2 RANGE=[Value]%
2
•
O2 RANGE=[Value]%
3
•
STABIL=[Value]PPM
4
•
COMEAS =[Value]MV
•
CO REF =[Value]MV
•
MR RATIO =[Value]
•
PRES =[Value]IN-HG-A
•
SAMP FL=[Value]CC/M
•
SAMPLE TEMP=[Value]ºC
•
BENCH TEMP=[Value]ºC
•
WHEEL TEMP=[Value]ºC
•
O2 CELL TEMP=[Value]ºC
3
•
BOX TEMP=[Value]ºC
•
PHT DRIVE =[Value]MV
•
SLOPE=[Value]
•
OFFSET=[Value]MV
•
CO2 SLOPE=[Value]
2
•
CO2 OFFSET=[Value]MV
2
•
O2 SLOPE=[Value]
3
•
O2 OFFSET=[Value]MV
3
•
TEST=[Value]MV
5
•
TIME=[HH:MM:SS]
Figure 4-2:
Viewing T300/T300M Test Functions
I
MPORTANT
I
MPACT ON
R
EADING OR
D
ATA
A value of “XXXX” displayed for any of the TEST functions indicates an
out-of-range reading or the analyzer’s inability to calculate it. All
pressure measurements are represented in terms of absolute pressure.
Absolute, atmospheric pressure is 29.92 in-Hg-A at sea level. It
decreases about 1 in-Hg per 300 m gain in altitude. A variety of factors
such as air conditioning and passing storms can cause changes in the
absolute atmospheric pressure.
06864D DCN7562