Teledyne API – Model T300/T300M CO Analyzer
Setup Menu
109
5.9.2.
ANALOG OUTPUT
Analog Output is used as a step test to check the accuracy and proper operation of the
analog outputs. The test forces all four analog output channels to produce signals
ranging from 0% to 100% of the full scale range in 20% increments. This test is useful
to verify the operation of the data logging/recording devices attached to the analyzer.
(See also Section
Access the Analog Output Step Test from the DIAG Menu as follows:
SAMPLE
RANGE=50.0 PPM CO= XXXX
<TST TST> CAL
SETUP
DIAG
SIGNAL I/O
PREV
NEXT
ENTR EXIT
DIAG AOUT
ANALOG OUTPUT
[20%]
EXIT
SETUP X.X
ENTER PASSWORD
8
1
8
ENTR
EXIT
DIAG AOUT
ANALOG OUTPUT
20%
EXIT
·
Pressing the button under “0%” pause
the test. Brackets will appear around the
value: EXAMPLE: [20%]
·
Pressing the same button again will
resume the test.
Performs analog output step
test 0% to 100%
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT