Teledyne API – Model T300/T300M CO Analyzer
Data Acquisition System (DAS) and APICOM
161
Triggering Events and Data Parameters/Functions for these default channels are:
Name:
STBSPN
Event:
EXITSP
Report Period: N/A
No. of Records:
200
RS-232 Report:
OFF
Channel Enabled:
ON
Cal Hold OFF:
OFF
Parameters:
2
DETMES
INST
1
OFF
RATIO
INST
3
OFF
Name:
CALDAT
Event:
SLPCHG
Report Period: N/A
No. of Records:
200
RS-232 Report:
OFF
Channel Enabled:
ON
Cal Hold OFF:
OFF
Parameters:
3
SLOPE1
INST
3
OFF
ZSCNC1
INST
1
OFF
OFFSET1
I NST
1
OFF
Name:
PNUMTC
Event:
ATIMER
Report Period: 000:01:00
No. of Records:
360
RS-232 Report:
OFF
Channel Enabled:
ON
Cal Hold OFF:
OFF
Parameters:
2
SMPLFLW
AVG
1
OFF
SMPLPRS
AVG
1
OFF
Name:
STBZERO
Event:
EXITZR
Report Period: N/A
No. of Records:
200
RS-232 Report:
OFF
Channel Enabled:
ON
Cal Hold OFF:
OFF
Parameters:
3
RATIO
INST
3
OFF
STABIL
INST
2
OFF
DETMES
INST
1
OFF
Name:
CONC
Event:
ATIMER
Report Period:
000:01:00
No. of Records:
800
RS-232 Report:
OFF
Channel Enabled:
ON
Cal Hold OFF:
ON
Parameters:
1
CONC1
AVG
1
OFF
PARAMETER
MODE
PRECISION
STORE NUM
SAMPLES
List of Parameters
List of Channels
Name:
TEMP
Event:
EXITSP
Report Period:
000:06:00
No. of Records:
400
RS-232 Report:
OFF
Channel Enabled:
ON
Cal Hold OFF:
OFF
Parameters:
3
PHTDRV
AVG
1
OFF
BNTEMP
AVG
1
OFF
BOXTEMP
AVG
1
OFF
Figure 7-1:
Default DAS Channel Setup