Teledyne API – Model T300/T300M CO Analyzer
Data Acquisition System (DAS) and APICOM
173
7.1.9.
RS-232 REPORT FUNCTION
The DAS can automatically report data to the communications ports, where they can be
captured with a terminal emulation program or simply viewed by the user using the
APICOM software.
To enable automatic COMM port reporting, follow the instruction shown in Section
7.1.4 then press:
Starting at the
EDIT CHANNEL MENU
SETUP X.X
0) CONC: ATIMER 1, 800
PREV MEXT
INS
DEL
EDIT
PRNT
EXIT
SETUP X.X
NAME: CONC
<SET
SET>
EDIT
EXIT
Continue pressing
SET>
until ...
SETUP X.X
RS-232 REPORT: OFF
<SET
SET
>
EDIT
EXIT
Use the
PREV
and
NEXT
buttons to
scroll to the
DATA
CHANNEL
to be
edited.
SETUP X.X
RS-232 REPORT: OFF
OFF
ENTR EXIT
EXIT
discards the new
setting.
ENTR
accepts the
new setting.
Toggle these buttons
to turn the
RS-232
REPORT
feature
ON/OFF.
7.1.9.1.
THE COMPACT REPORT FEATURE
When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead
of reporting each parameter in one channel on a separate line, up to five parameters are
reported in one line.
The
COMPACT DATA REPORT
generally cannot be accessed from the standard
DAS front panel menu, but is available via the instrument’s communication ports by
using APICOM or the analyzer’s standard serial data protocol.