Teledyne API – Model T300/T300M CO Analyzer
Getting Started
75
S
ET
CO
S
PAN
G
AS
C
ONCENTRATION
Set the expected CO pan gas concentration. This should be 80-90% of range of
concentration range for which the analyzer’s analog output range is set.
M-P CAL CO SPAN CONC:40.0 Conc
0
0
4
0
.0
0
ENTR EXIT
SAMPLE
GAS TO CAL:CO
CO
O2
ENTR
EXIT
SAMPLE RANGE=50.0 PPM CO= XX.XX
<TST TST>
CAL
SETUP
M-P CAL RANGE=50.0 PPM CO= XX.XX
<TST TST> ZERO SPAN
CONC
EXIT
SAMPLE
RANGE TO CAL:LOW
LOW
HIGH
ENTR
EXIT
The CO span concentration value is
automatically default to
40.0 PPM.
If this is not the the concentration of
the span gas being used, toggle
these buttons to set the correct
concentration of the CO
calibration gas.
EXIT ignores the new
setting and returns to
the previous display.
ENTR accepts the new
setting and returns to
the
CONCENTRATION
MENU.
Only appears if either
the optional O2 or CO2
sensors are installed.
Only appears if the
analyzer is set for DUAL
range (LOW/HIGH) mode
06864D DCN7562