Getting Started
Teledyne API – Model T300/T300M CO Analyzer
76
Z
ERO
/S
PAN
C
ALIBRATION
To perform the zero/span calibration procedure, press:
Press
ENTR
to changes
the
OFFSET
&
SLOPE
values for the CO
measurements.
Press
EXIT
to leave the
calibration unchanged and
return to the previous
menu.
SAMPLE
STABIL= XXXX PPM
CO=XX.XX
< TST TST > CAL SETUP
M-P CAL
STABIL= XXXX PPM
CO=XX.XX
<TST TST>
ZERO
CONC
EXIT
EXIT
at this point
returns to the
SAMPLE menu.
Press
ENTR
to changes
the
OFFSET
&
SLOPE
values for the CO
measurements.
Press
EXIT
to leave the
calibration unchanged and
return to the previous
menu.
SAMPLE RANGE=0.0 PPm CO= XX.XX
< TST
TST >
CAL SETUP
Set the Display to show
the
STABIL
test function.
This function calculates
the stability of the CO
measurement.
Toggle
TST>
button until ...
Allow zero gas to enter the sample port
at the rear of the analyzer.
Wait until
STABIL
falls below
0.5 ppm.
This may take several
minutes.
SAMPLE
STABIL= XXXX PPM
CO=XX.XX
< TST TST >
CAL
SETUP
M-P CAL
STABIL= XXXX PPM
CO=XX.XX
<TST TST>
ENTR
CONC
EXIT
M-P CAL
STABIL= XXXX PPM
CO=XX.XX
<TST TST> ZERO
SPAN
CONC
EXIT
Allow span gas to enter the sample port
at the rear of the analyzer.
Wait until
STABIL
falls below
0.5 PPM.
This may take several
minutes.
SAMPLE
STABIL= XXXX PPM
CO=XX.XX
< TST TST >
CAL
SETUP
M-P CAL
STABIL= XXXX PPM
CO=XX.XX
<TST TST>
ENTR
CONC
EXIT
M-P CAL
STABIL= XXXX PPM
CO=XX.XX
<TST TST> ENTR
CONC
EXIT
The
SPAN
button now
appears during the transition
from zero to span.
You may see both keys.
If either the
ZERO
or
SPAN
buttons fail to appear see the
Troubleshooting section for
tips.
Figure 3-29:
Zero/Span Calibration Procedure
06864D DCN7562