12
Teledyne API T500U CAPS NO
2
Analyzer with NumaView™ Software
083730510A DCN8172
Parameter
Description
Sample Flow Rate
900 cm
3
/min
±
10%
AC Power
Rating
Typical Power Consumption
110 - 120 V~ 60 Hz 3.0 A
220 - 240 V~ 50 Hz 3.0 A
80 W
80 W
Analog Output Ranges
10V, 5V, 1V, 0.1V (selectable)
Analog Output Resolution 1 part in 4096 of selected full-scale voltage
Communications
Standard I/O 1 Ethernet: 10/100Base-T
2 RS-232 (300 – 115,200 baud)
2 USB device ports
8 opto-isolated digital outputs
6 opto-isolated digital inputs (3 defined, 3 spare)
4 analog outputs
Optional I/O 1 USB com port
1 RS485
4 digital alarm outputs
Multidrop RS232
2 4-20mA current outputs
Operating Temperature
5-40
o
C
Humidity Range
0-95% RH, Non-Condensing
Dimensions HxWxD
7” x 17” x 23.5” (178 x 432 x 597 mm)
(19” rack mount, 5U, 24” deep)
Weight
33 lbs (15 kg)
Environmental Conditions
•
Installation Category (Over Voltage Category ) II Pollution
Degree 2
•
Intended for Indoor Use Only at Altitudes
≤
2000m
Note: All specifications are based on constant conditions
EPA DESIGNATION
Teledyne API’s Model T500U cavity attenuated phase shift spectroscopy nitrogen dioxide
analyzer is officially designated as a US EPA Federal Equivalent Method (FEM),
Designation number EQNA-0514-212 for NO
2
measurement as defined in 40 CFR Part 53.
The official List of Designated Reference and Equivalent Methods is published in the U.S.
Federal Register
–
http://www3.epa.gov/ttn/amtic/criteria.html
and specifies the
instrument’s settings and configurations.
SAFETY
IEC/EN 61010-1:2010 (3
rd
Edition), Safety requirements for electrical equipment for
measurement, control, and laboratory use.
CE
:
2006/95/EC, Low-Voltage Directive
Summary of Contents for T500U
Page 2: ......