CONNECTING ANALOG OUTPUTS
The rear panel Analog Output channels A1 through A4 can be mapped to reflect various
operating values in the analyzer, including concentration values, temperatures, pressures,
etc. These mappings are not configured by default and must be set by the user.
An optional Current Loop output (Section 2.3.1.3) is available for A1, A2 and A3 only.
To access these signals attach a strip chart recorder and/or data-logger to the appropriate
analog output connections on the rear panel of the analyzer.
Configure through the Setup>Analog Outputs menu (Section 2.5.8)
ANALOG OUT
A1 A2 A3 A4
+ - + - + - + -
Figure 2-4. Analog Output Connector
Table 2-3. Analog Output Pin Assignments
PIN
ANALOG
OUTPUT
SIGNAL OUT
STANDARD
VOLTAGE OUTPUT
CURRENT
LOOP OPTION
1
A1
User-selected
through the
Setup>Analog
Outputs menu.
V +
I Out +
2
Ground
I Out -
3
A2
V +
I Out +
4
Ground
I Out -
3
A3
V +
I Out +
4
Ground
I Out -
7
A4
V +
Not Available
8
Ground
Not Available