24
Teledyne API T500U CAPS NO
2
Analyzer with NumaView™ Software
083730510A DCN8172
Table 2-4. Status Output Pin Assignments
PIN
STATUS
DEFINITION
CONDITION
1-8
Configurable
through the
Setup>Digital
Outputs menu
Collector side of individual status output opto-isolators.
D
Emitter BUS
The emitters of the transistors on pins 1 to 8 are
bussed together.
Blank
NO CONNECTION
+
DC Power
+ 5 VDC, 300 mA source maximum
Digital Ground
The ground level from the analyzer’s internal DC power
supplies. This connection should be used as the
ground return when +5VDC power is used.
CONNECTING THE CONTROL INPUTS (DIGITAL INPUTS)
With zero and span valves option installed, two digital control inputs in the rear panel
CONTROL IN connector, can be used to remotely activate the zero and span calibration
modes (see Section 4.2.2.1).
Energize the Control Inputs either by the in5V available from the pin labeled “+”
(more convenient), or by a separate external 5 VDC power supply (ensures that these inputs
are truly isolated). Refer to Figure 2-7.
CONTROL IN
A B C D E F U
+
S
PA
N
Z
E
R
O
CONTROL IN
A B C D E F U
+
S
PA
N
Z
E
R
O
-
+
5 VDC Power
Supply
Local Power Connections
External Power Connections
Figure 2-7. Energizing the Control Inputs