Alerts are notifications triggered by specific criteria having been met by either factory-
defined conditions (standard and not editable) or user-defined Events (Section 2.5.2). The
Active Alerts page shows the status of any active warning conditions or Events that have
been triggered.
When Alerts are triggered, a caution symbol appears in both the Alerts menu tab and in the
bottom right corner of the software interface, which serves as a shortcut to the Alerts page
from any other page. View a list of currently active Alerts by pressing either the Alerts
menu on the Home screen or by pressing the Alerts shortcut (Figure 2-23).
While Alerts can be cleared from the Active Alerts page, they remain recorded in the
Utilities>Alerts Log menu.
Figure 2-23. Navigating to the Active Alerts Page
Alerts can be configured as either latching (appears in Active Alerts screen when Event is
triggered and must be cleared by the user) or non-latching (Active Alerts screen
continuously updates based on the Event criteria, clearing on its own). See Section 2.5.2.
To clear Alerts from the Active Alerts page, either check individual boxes to choose
specific Alerts, or check the Select All box to choose all Alerts, then press the Clear
Selected button.