083730510A DCN8172
Teledyne API T500U CAPS NO
2
Analyzer with NumaView™ Software
57
S
ETUP
>A
NALOG
O
UTPUTS
Map the four user-configurable Analog Outputs to any of a wide variety of “Signals”
present in the instrument and customize their respective configurations.
Figure 2-39. Analog Output Configuration Example
Refer to Figure 2-39 for the following:
•
Signal Out: select a Signal for the output.
•
Min/Max: edit minimum and maximum values associated with the selected Signal.
•
Calibration Type:
•
AUTO for group calibration (Figure 2-40) of the analog outputs (cannot be
selected when Current is selected for the Range)
•
MANUAL for individual calibration (Figure 2-41) of analog outputs where manual
adjustments can be made (the only calibration type allowed when Current is
selected for the Range). See Sections 2.5.8.1 and 2.5.8.2 .
•
Range: assign a voltage or select Current (refer to Table 2-9).
•
Recorder Offset: add a zero offset for recording slightly negative readings from noise
around the zero point.
•
Allow Overrange: check to allow a ± 5% over-range; uncheck to disable over-range if
the recording device is sensitive to excess voltage or current.
Summary of Contents for T500U
Page 2: ......