Figure 2-44.
Alternative Setup Using 250Ω Resistor for Checking Current Output Signal Levels
In this case, follow the procedure above but adjust the output for the following values:
Table 2-11. Current Loop Output Check
% FS
Voltage across Resistor for 2-20 mA
Voltage across Resistor for 4-20 mA
0
500 mVDC
1000 mVDC
100
5000 mVDC
5000 mVDC
S
ETUP
>I
NSTRUMENT
As presented in Table 2-12, view product and system information and network settings,
edit network settings, and perform certain maintenance tasks.
Table 2-12. Setup>Instrument Menu
MENU
DESCRIPTION
Product Info
View Model, Part, and Serial Numbers and Package and Driver
Versions, and options information.
System Info
View Windows and RAM information.
Network Settings
View the network settings (configurable through the
Setup>Comm>Network Settings menu).
Date/Time Settings
Adjust date, hour, and minutes, select a time zone*, and set the system
clock to automatically adjust for Daylight Savings Time or not. (Also see
Setup>Vars>Daylight Savings Enable). *Time Zone change requires a
special procedure; see Maintenance Section 5.5.
NTP Time Settings
Configure Network Time Protocol settings for clock synchronization.
Language
Select an available language.
Remote Update
When an instrument is connected to a network that is connected to the
Internet, follow the instructions on this Remote Update page to check for
and activate software/firmware updates. (Also refer to Section 5.3).