Table 3-3. Hessen Status Flags and Default Bit Assignments
STATUS FLAG NAME
DEFAULT BIT ASSIGNMENT
HEX BITS
OPERATIONAL FLAGS (8 BITS)
In MANUAL Calibration Mode
x02
0000 0010
In ZERO Calibration Mode
x04
0000 0100
In SPAN Calibration Mode
x08
0000 1000
INVALID CONC
x80
1000 0000
SPARE/UNUSED
x01
0000 0001
UNITS OF MEASURE FLAGS (ONLY BITS 5 AND 6 OF THE OPERATIONAL
FLAG)
UGM
00
MGM
01
PPB
10
PPM
11
WARNING FLAGS (8 BITS)
ANY PRESS WARN
x04
0000 0100
SAMPLE TEMP WARNING
x08
0000 1000
INTERNAL PUMP OVERRIDE SET TO OFF
x10
0001 0000
IZS TEMP WARNING
1
x20
0010 0000
BENCH LED DISABLED
x40
0100 0000
OVEN TEMP WARNING
x80
1000 0000
UNASSIGNED FLAGS
MANIFOLD TEMPERATURE
2
FRONT PANEL COMMUNICATION WARNING
BOX TEMP WARNING
ANALOG CALIBRATION WARNING
SYSTEM RESET
DYNAMIC ZERO WARNING
RELAY BOARD WARNING
DYNAMIC SPAN WARNING
REAR BOARD NOT DETECTED
IN MULTI-POINT CALIBRATION MODE
AUTOREF WARNING
1
Only applicable if the optional internal span gas generator is installed.
2
Only applicable if the T500U is equipped with the optional manifold.
It is possible to assign more than one flag to the same Hessen status bit. This allows the grouping of
similar flags, such as all temperature warnings, under the same status bit.
Be careful not to assign conflicting flags to the same bit as each status bit will be triggered if any of the
assigned flags is active.
HESSEN GAS LIST
In the T500U there is no Hessen Gas List to configure as there is only one gas, NO2, the
Hessen ID for which is 213.