82
Teledyne API T500U CAPS NO
2
Analyzer with NumaView™ Software
083730510A DCN8172
EPA PROTOCOL CALIBRATION
When running this instrument for U.S. EPA compliance, always calibrate prior to use,
adhering to the EPA designation requirements for this instrument. (The official List of
Designated Reference and Equivalent Methods is published in the U.S. Federal Register:
http://www3.epa.gov/ttn/amtic/criteria.html
this List specifies the settings and
configurations for EPA calibration protocol). Pay strict attention to the built-in warning
features, periodic inspection, regular zero/span checks, regular test parameter evaluation
for predictive diagnostics and data analysis, and routine maintenance. Any instrument(s)
supplying the zero air and span calibration gasses used must themselves be calibrated, and
that calibration must be traceable to an EPA/NIST primary standard.
Comply with Code of Federal Regulations, Title 40 (downloadable from the U.S.
Government Publishing Office at
) and with Quality Assurance
Guidance documents (available on the EPA website:
http://www3.epa.gov/ttn/amtic/qalist.html
). Give special attention to specific regulations
regarding the use and operation of ambient NO
X
analyzers (chemiluminescence).
Summary of Contents for T500U
Page 2: ......