on the probe position, and could be larger if the transmitter is positioned close to the DUT
ground. Note that the HVFO uses a dual-reference lead (as described earlier) that separates the
common-mode current from the signal return current. This greatly reduces the effect of the
ground current due to the common-mode capacitance and results in CMRR performance that is
much better than would otherwise be expected for an ~7 pF common-mode capacitance.
Compatibility
The HVFO108 probe is compatible with a wide range of Teledyne LeCroy oscilloscopes equipped
with the ProBus interface. Visit teledynelecroy.com/oscilloscope to confirm compatibility with
your oscilloscope model.
In general, Teledyne LeCroy 12-bit High Definition Oscilloscopes (HDO) provide the most offset
capability over the widest range of V/div settings.
Note:
HVFO108 is not compatible with WaveSurfer Xs oscilloscopes, which only run up to 32-bit
XstreamDSO 7.2.0.5. Contact Service for available oscilloscope upgrades.
Required Firmware Versions
Correct operation of the HVFO108 probe requires that the oscilloscope be installed with
firmware version 9.2.x.x. or higher.