HVFO108 High-Voltage Fiber Optically Isolated Probe
12
EAR99 technology subject to restrictions on copyrights page.
HVFO108 Probe Specifications
For the current specifications, see the product datasheet at teledynelecroy.com. Specifications
are subject to change without notice.
Performance
Bandwidth (probe + oscilloscope)
150 MHz typical
Risetime 10-90%
3.3 ns typical
Gain Accuracy
+/- 2.5% typical, 4% guaranteed
CMRR (typical, 1x tip attached)
140 dB @ 100 Hz
120 dB @ 100 Hz - 1 MHz
85 dB @ 1 - 10 MHz
60 dB @ 10 - 60 MHz
35 dB @ 150 MHz
Electrical Characteristics
Input Dynamic Voltage Range
1x tip
5x tip
10x tip
20x tip
40x tip
+/- 1 V
+/- 5 V
+/- 10 V
+/- 20 V
+/- 40 V
Max. Common Mode Voltage
(from either input to ground)
+/- 35 kV (DC + peak AC)
Max. Input Voltage to Earth
(from either input to ground)
+/- 35 kV (DC + peak AC)
Max. Safe Input Voltage
30 V
RMS
/ 60 V
DC
(hand-held)
Max. Non-destruct Voltage
Lower of 5X dynamic range
or 100 V
Noise (probe only)
1x tip
5x tip
10x tip
20x tip
40x tip
7 mVrms
35 mVrms
70 mVrms
140 mVrms
280 mVrms
WARNING
: The probe is not intended for hand-held use above 60 V
DC
. Always maintain
adequate spacing between floating probe components and earth ground when
measuring at high voltage in a stationary test setup.